High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection

Labnovation Technologies, Inc.
Contact Now

High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection

Brand Name : LABNOVATION

Model Number : LD-560

Certification : IFCC, NGSP, CE

Place of Origin : Shenzhen, China

MOQ : 1 Piece

Price : US$5000-12000/Piece

Payment Terms : D/A, D/P, T/T

Supply Ability : 100 Units/Day

Delivery Time : 5-7 work day

Packaging Details : Carton

Method : HPLC (High Performance Liquid Chromatography))

CertificatedWaveLength : 420nm

Sample Capacity : 20 Samples

Parameters : HbA1c, HbF, HbA1a, HbA1b, HbA0; HbA2 (optional)

Certificate : NGSP, IFCC, CE

Working Condition : Temperature: 15-30℃ Humidity:30%-85%

Voltage : 100-240VA

Report Range : 3-18%

Contact Now

LD-560 HPLC System for HbA1c Testing

The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin

(HbA1c) in human blood.

High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection

Performance Parameteras


Reporting unit: Glycosylated hemoglobin (HbA1c) is reported in NGSP (%) and IFCC units (mmol/mol).
Accuracy: The accuracy is tested with reference materials as samples, and the relative deviation of the
test results should be within ± 5%.
Repeatability: For samples with a test concentration of 4% 6%, the sample coefficient of variation (CV)
of the test results is not greater than 2.0%.
Linear range: Within the range of 3% 18% of glycosylated hemoglobin, the linear correlation coefficien t
R value of the test results should not be less than 0.9900.
Stability: Within 8 hours after the analyzer is turned on and stabilized, the same normal sample is tested, and the relative deviation of the test results is not more than ±3.0%.
Carrying pollution: The carrying pollution rate of the analyzer should not be greater than 2.0%.
Basic function: Provides Chinese report and abnormal alarm.

The temperature accuracy and fluctuation of the chromatography column module: the temperature is within ±1°C of the set value, and the fluctuation is not more than 0.8°C.

Certificate

High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection

High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection


China High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection wholesale

High Performance Liquid Chromatography Automated HbA1c Analyzer For Glycosylation Hemoglobin Detection Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Labnovation Technologies, Inc.
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0